P472 - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
P472 - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany

P472 – Exit Slit Unit STXM EMIL BESSY

Application

Exit slit unit for soft X-ray synchrotron radiation

Year of delivery

2019

Installation site

STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany

Design Features

  • Double slit mechanism – vertically and horizontally defining slits.

  • Motorized and encoded slits opening, symmetric opening, non-linear opening characteristics.

  • OFHC slit blades.
  • Electrical isolation of horizontally defining slit blades.
  • Motorized and encoded translation of the horizontally defining slit transversal to the beam axis.
  • Manual rotation of the vertically defining slit mechanism around the beam axis.
  • Motorized and encoded slit unit translation along beam axis.
  • Isolated bias plate.
  • Adjustable support structure.

Performance Features

  • Precision and slit opening repeatability smaller than 100 nm.
  • Parallelism smaller than 1 µm over 10 mm edge length.
  • Straightness smaller than 0.25 µm over 10 mm edgle length.
  • Very small deviation (submicrometer) of the slit opening calibration curve from the theoretical non-linear slit opening function.

Outer Dimensions

P472 - Outer dimensions - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany

Technical specifications and performance values

General

Vacuum chamber

Chamber type

cylindrical, horizontal axis

Main flanges

DN160CF cluster flanges

Material

1.4301

Base pressure

10-10 mbar range

Translation along beam axis

Range

± 250 mm

Resolution (design value)

0.025 µm /motor fine step
(1/16, recommended)
0.01 µm /encoder count

Horizontally defining slit (HDS)

Slit blades

Material

Hardened stainless steel, gold coated

Straightness

< 0.20 µm over 10 mm (1σ)

Parallelism

-0.03 µm /mm
(20 µm slit opening)
0.01 µm /mm
(50 µm slit opening)
0.003 µm /mm
(100 µm slit opening)

Slit opening

Range

0 µm to 6 mm

Repeatability

± 0.033 µm at 20 µm slit opening
± 0.023 µm at 50 µm slit opening
± 0.009 µm at 100 µm slit opening

Stability (> 30 min)

0.002µm

Resolution (design value)

0.006 µm /motor fine step
(1/16, recommended)
0.01 µm /encoder count

Slit translation

Range

-5.4 mm to 6 mm

Resolution (design value)

0.006 µm /motor fine step
(1/16, recommended)
0.01 µm /encoder count

Vertically defining slit (VDS)

Slit blades

Material

Hardened stainless steel

Straightness

< 0.25 µm over 10 mm (1σ)

Parallelism

-0.07 µm /mm
(20 µm slit opening)
0.06 µm /mm
(50 µm slit opening)
0.002 µm /mm
(up to 100 µm slit opening)

Slit opening

Range

0 µm to 2.3 mm

Repeatability

± 0.011 µm at 10 µm slit opening
± 0.002 µm at 50 µm slit opening
± 0.003 µm at 100 µm slit opening

Long-term stability (> 2 h)

0.004 µm

Resolution (design value)

0.006 µm /motor fine step
(1/16, recommended)
0.01 µm /encoder count

Slit rotation

Range

– 0.5° to 0.5°

Actuator

manual rotary feedthrough

Resolution  (design value)

0.153° /turn

Performance Test Results

Slit width calibration (HDS)
P472 - Slit width calibration (HDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit opening repeatability (HDS)
P472 - Slit opening repeatability (HDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit blades parallelism (HDS)
P472 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit blades parallelism (HDS)
P472 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit blades parallelism (HDS)
P472 - Slit blades parallelism (HDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit opening stability (HDS)
P472 - Slit opening stability (HDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit width calibration (VDS)
P472 - Slit width calibration (VDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit opening repeatability (VDS)
P472 - Slit opening repeatability (VDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit blades parallelism (VDS)
P472 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit blades parallelism (VDS)
P472 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit blades parallelism (VDS)
P472 - Slit blades parallelism (VDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany
Slit opening long-term stability (VDS)
P472 - Slit opening stability (VDS) - Exit slit unit for soft X-ray synchrotron radiation STXM beamline at BESSY, Helmholtz-Zentrum Berlin, Germany